General Information:
EPOL60+ Automatic Oil Dielectric Loss and Resistivity Tester, is used for measuring dielectric dissipation factor and DC resistivity of insulating oil and other insulating liquids. The instrument integrates with test oil cell, temperature controller, temperature sensor, dielectric loss test bridge, AC test power, standard capacitor, high resistance meter, DC high voltage source and other major components.
Features:
-The Oil vessel adopts the three-electrode structure according to standard IEC 60247:2004, with electrode spacing 2mm, which can eliminate the effects of the stray capacitance and leakage on dielectric loss test results.
-The instrument adopts the medium frequency induction heating, PID temperature control algorithm. This heating method has the advantages of non-contact between the test cell and the heating body, uniform heating, fast speed, convenient control, and strictly control of the temperature within error range of the preset temperature.
-Internal standard capacitor is SF6 gas-filled three-electrode capacitance, and its capacitor dielectric loss and electrical capacity are not affected by ambient temperature, humidity etc, so that the instrument accuracy can be ensured even after prolonged use.
-AC test power supply uses AC-DC-AC conversion, which effectively prevents the fluctuations in mains voltage and frequency from affecting the dielectric loss measurement accuracy. Even use a generator, the instrument can work properly.
-Perfect protection function. When there is over voltage, over current or high-voltage short circuit, instruments can quickly cut off high voltage, and issues a warning message. When the temperature of the sensor failure or not connected, the instrument will also issue a warning message.
-There is a temperature limiting relay in the medium frequency induction heating furnace. When temperature above 120 ℃, the instrument frees the relay and stops heating.
-Convenient test parameter setting. Temperature setting range 0~120ºC, AC voltage setting range 200~2200V, DC voltage setting range 0~500V.
-Large TFT touch screen, with backlight and clear display. With simply set up, the instrument can be automatically tested. And automatically store and print test results.
-Built-in real-time clock, the test date and time can be saved, displayed and printed with the test results.
-Calibration functions of cleaned dry vessel. The measurement of dry vessel’s capacitance and dielectric loss factor can determine its cleaning and assembly conditions. The calibration data is automatically saved and so that it facilitates the calculation of relative permittivity and DC resistivity.
-The instrument takes the RAM9 platform as the core, the test accuracy is high and the speed is fast.
Technical Specifications:
Voltage supply: | AC 220V±10% |
Power frequency: | 50Hz/60Hz ±1% |
Measuring range | |
Capacitance: | 5pF~200pF |
Relative Permittivity: | 1.000~30.000 |
Dielectric loss factor: | 0.00001~100 |
DC resistivity: | 2.5 MΩm~20 TΩm |
Measuring Accuracy | |
Capacitance: | ±(1%reading+0.5pF) |
Relative permittivity: | ±1% of reading |
Dielectric dissipation factor: | ±(1%reading+0.0001) |
DC resistivity: | ±10% of reading |
Resolution | |
Capacitance: | 0.01pF |
Relative capacitively: | 0.001 |
Dielectric dissipation factor: | 0.00001 |
Measured temperature range: | 40~120ºC |
Temperature measurement error: | ±0.5ºC |
AC Test Voltage: | 200~2200V Continuously adjustable, frequency 50Hz |
Power consumption: | 100W |
Ambient temperature: | 0ºC~40ºC |
Relative humidity (RH): | <75% |
Dimensions: | 460mm×370mm×330mm |
Total weight: | 25KG |