General Information:
EPR50C Digital insulation resistance meter also known as megohmmeter, high voltage insulation resistance tester, etc., for the insulation resistance test. The instrument has a large LCD screen gray backlit display, data storage, data access, alarm, automatic shutdown and other functions. At the same time, it also has the function of measuring AC voltage absorption ratio and polarization index of DC voltage. The machine is beautiful and upscale, has a wide range, high resolution, convenient operation, easy to carry, accurate, reliable, stable performance, strong anti-interference ability.
EPR50C has a shockproof, dustproof, moisture-proof structure and is a commonly used and indispensable instrument for telecommunications, electricity, meteorology, computer rooms, oil fields, electromechanical installation and maintenance, and industrial enterprises that use electricity as industrial power or energy. It is suitable for measuring the resistance value of various insulating materials and the insulation resistance of transformers, motors, cables and electrical equipment.
EPR50C Digital insulation resistance meter consists of medium and large scale integrated circuits. The output power of the watch is large, and the output voltage level is a lot (with 6 voltage levels). Rated insulation test voltage range 250V ~ 10kV, insulation resistance measurement range 0.01MΩ ~ 2000GΩ. DC voltage measurement range 0 ~ 1000V, AC voltage measurement range 0 ~ 750V.
Measuring Range and Accuracy:
Test Function |
Insulation resistance test, voltage test |
|
Condition |
23°C±5°C,below 75%rh |
|
Rated voltage (V) |
250, 500, 1000, 2500V, 5000V, 10000V |
|
Measuring voltage(V) |
Rated voltage×(1±10%) |
|
Insulation Resistance Range(GΩ) |
0.01MΩ~2000GΩ Precision ±3%rdg±5dgt |
|
Insulation Resistance Resolution |
0.01MΩ |
|
DC voltage range |
0~1000V |
Resolution 0.1V |
AC voltage range |
0~750V |
Resolution 0.1V |
Output short circuit current |
≥5mA(10000V) |
|
Absorption Ratio and Polarization Index Measurement |
Have |
|
Power Supply |
12V rechargeable battery |
|
Backlight |
Controllable gray screen backlight, suitable for use in dim places |
|
Display mode |
4-bit large LCD display, gray screen backlight |
|
LCD display size |
108mm×65mm |
|
Instrument size |
227.2mm (L)×227.5mm (W)×153mm (H) |
|
USB interface |
With USB interface, software monitoring, storage data can be uploaded to the computer, save the print |
|
Communication Line |
USB communication line 1PC |
|
Test Line |
High voltage robs red 3meter, high voltage test line black 1.5meter, green 1.5meter |
|
Data Storage |
500 groups, "FULL" symbol indicates that storage is full |
|
Data Review |
Data review function: "READ" symbol display |
|
Overflow Display |
Exceed measurement range overflow function: “OL” symbol display |
|
Alarm Function |
Alarm when the measured value exceeds the alarm setting value |
|
Power Consumption |
Standby: 30mA Max (backlight off) |
|
Turn on backlight: 42mA Max |
||
Measure:300mA Max (backlight off) |
||
Instrument weight |
2720g (including battery) |
|
Battery voltage |
|
|
Automatic shut-down |
automatic shutdown after 15 minutes boot |
|
Insulation Resistance |
≥50MΩ (between Measuring line and shell) |
|
Pressure Resistance |
AC3kV/50Hz 1min |
|
Working Temperature and Humidity |
-10℃~+50℃<85%RH |
|
Storage Temperature and Humidity |
-15℃~+55℃<90%RH |
|
Suitable for Safety Regulations |
IEC61010-1, IEC1010-2-31, IEC61557-1,5, IEC60529(IP54), Pollution grade 2, CAT III 300V |