General Information:
Automatic Closed Cup Oil Flash Point Tester EPT404, mainly used for the determination of closed flash point value of petroleum products. The instrument adopts ARM microprocessor technology, color LCD, resistive touch screen technology, and English menu, which is more convenient for human-computer interaction; the instrument has power-off storage Functions: The instrument has the functions of automatic ignition, display, locking and printing results, and automatic cooling; the instrument has the advantages of accurate measurement, good repeatability, stable and reliable performance, and simple operation. It is widely used in electric power, petroleum, chemical, commodity inspection, scientific research and other departments, in line with ASTM D93, GB/T 261-2008 standard method requirements.
Technical Specifications:
Color LCD display |
|
Operation mode |
Touch screen |
Measuring range |
40~370℃ |
Temperature detection |
Platinum resistance |
Accuracy |
≥110℃±2℃ ≤110℃±1℃ |
Ignition method |
Electric ignition |
Information storage |
1000 measurement results can be stored |
Cooling method |
Forced air cooling |
Printer |
Thermal |
Self-check function |
Test arm, igniter, printer, etc. |
Repeatability |
ASTM D93 GB/T 261 standard |
Power supply |
AC 220V±11V, 50Hz±2.5HZ |
Power |
≤350VA |
Use environment temperature |
10℃~35℃ |
Use environment humidity |
≤85% |